is an open system that provides flexible data analysis and storage.
A non-volatile memory stores up to 120 user-programmed tests.
The Auto-Track™ transport system can load up to three strips (up to 36 wells total) and handles break-apart, single, or double strips.
The "Home Sensor" requires careful reassembly of the PCA connections (J30, J31, J32, and J33) to prevent malfunction errors. Troubleshooting Common Errors
Features a linear measurement range of 0.0 to 3.0 Absorbance (A).
Accessing internal components requires inverting the instrument and removing the enclosure screws using a Philips #1 screwdriver.
A common technical service procedure involves replacing the diffuser tape on the guide bracket aperture to maintain consistent light distribution.